MSL Cable Termination
114
Installin
the Instrument
Step 5 Slide the PCB back on to the metallic mounting flange.
Step 6 Use screws to fasten the mounting flange to the wall.
NOTE US version only: Fasten the rectangular wall-mounting plate to the wall. Attach the mounting
flange to the wall-mounting plate.
Step 7 Mount the plastic cover. The plastic cover consists of two pieces:
Frame
Angled cover
Put the frame over the mounting insert and the PCB. Place the angled cover on top
and fasten with two screws.
Step 8 Connect the monitor and the measurement server to the wall installation.
Step 9 Perform the following tests as described in the “Testing the Instrument” section of
this manual:
Power-on test blocks
Safety test blocks
ECG Sync Performance Test