Part 2 Isolating and Solving Instrument Problems
Troubleshootin
g
the Instrument
189
Display waves
are noisy/not
showing a
smooth QRS
wave.
Radio frequency interfer-
ence (RFI) from other
instruments such as ultra-
sound, ESU, Defib, or
telemetry is causing noisy
waveforms.
Turn off or move away
from the Monitor the
instrument causing RFI
problems.
Message
appears: New
display
detected, please
enter new dis-
play specs”
The display has been
exchanged and the Moni-
tor needs the new specifi-
cations.
Use the Upgrade Tool to
upgrade the monitor with
the latest display specifi-
cations.
TouchStrip related problems
TouchStrip dis-
colored, cracked
or surface is
deformed.
Cleaning agents, abrasive
material, or strong sol-
vents were used to clean
the Instrument.
Replace the Monitor
Bezel and clean the Instru-
ment in the future with
only isopropyl alcohol.
Wipe cleaning agents off
surfaces immediately.
Lack of user
control of dis-
play.
Cable from the System
Board to the Monitor
Bezel not connected prop-
erly.
Defective TouchStrip.
Defective System Board.
Refit cable correctly.
Check Status Log for mes-
sage: CHECK INPUT
DEVICE
Replace the System
Board.
Network related problems
“No Central
assigned to this
bed” prompt
message
Wrong monitor label or
Information Center incor-
rectly configured
Check the bed assign-
ments at the Information
Center and that the moni-
tor label assigned to this
monitor has not been
changed locally