Part 2 Isolating and Solving Instrument Problems
190
Troubleshootin
the Instrument
“No Central -
duplicate moni-
tor label”
prompt message
Monitor label configura-
tion is in conflict with
another monitor
Check that the monitor
label assigned to this mon-
itor has not been changed
locally. Check the assign-
ment at the Information
Center
“Assigned Cen-
tral is not avail-
able” prompt
message
The Information Center is
switched off or not acces-
sible
“No Central -
software incom-
patible” prompt
message
Ask Biomed department
to check software revi-
sions.
“Central can not
identify this
bed” prompt
message
The location of this bed is
not clear.
Check the bed assign-
ments at the Information
Center.
Measurement Server related problems
No waves. Measurement Server
defective.
LEDs on display unit
blink but Server LED does
not.
Check green LED on
Server. Exchange the
Server.
Connection between the
Server and the display is
lost. Exchange the Server.
No ECG data or
waveform dis-
played when
both defib sync
and balloon
pump are done
together.
2-circuit cable used in a 3-
circuit ECG output jack.
Use a bud box when set-
ting up ECG.