Part 2 Isolating and Solving Instrument Problems
Troubleshootin
the Instrument
191
Red ECG LED
is on.
System Board is defec-
tive.
Defib. synchronization is
not in sync. with ECG.
Defib. synchronization
fails.
Replace the System
Board.
Exchange the Server and
try again. Check the defib.
cable.
SpO
2
values
appear unstable
or unreliable
Low light transmission of
SpO
2
sensor LEDs to the
sensor photo diode.
Patient has low peripheral
perfusion
Check Transmission and
Perfusion Index as fol-
lows:
1) Fit SpO
2
sensor on your
finger
2) Enter Config Mode
3) Select SpO
2
.
4) Switch SpO
2
On
Lightly press the Touch-
strip until text appears at
the bottom of the screen:
Perfusion Index xx.x
Transm. red yyyy
Transm. infra red zzzz
The Perfusion Index
should be > 1 and both
Transmission indexes
should be greater than 10.