3D s / 3D Classic / 3D Plus or 3D Mid with X-ray
tube D-054SB or 3D Max with X-ray tube
D-067SB
3D: 1 - 14 mA ±10%
Pan / SmartPan: 1 - 16 mA ±10%
Scanning ceph: 1 - 16 mA ±10%
Planmeca ProCeph: 16 mA ±10%
3D Mid / 3D Max with X-ray tube D-059SBR
3D: 1 - 14 mA ±10%
Pan / SmartPan: 1 - 14 mA ±10%
Scanning ceph: 1 - 14 mA ±10%
Planmeca ProCeph: 11 mA ±10%
3D Mid / 3D Max with X-ray tube SXR
130-10-0.5 SC
3D: 1 - 14 mA ±10%
Pan / SmartPan: 1 - 16 mA ±10%
Scanning ceph: 1 - 16 mA ±10%
Planmeca ProCeph: 11 mA ±10%
mAs range
min. / max. as indicated ±(10% + 0.2 mAs)
mGy range
min. / max. as indicated ±40%
Linearity of radiation output
< 0.1
DEC accuracy
±10%
Cooling period
Automatically controlled
Exposure time
Pan 2.7 - 16 s as indicated ±10%
SmartPan 2.5 - 15.6 s as indicated ±10%
Scanning ceph 6.7 - 10.5 s as indicated ±10%
Planmeca ProCeph 0.1 - 0.8 s as indicated ±10%
3D Pulsed, effective 3 - 36 s as indicated ±10%
SID
Pan 2D / 3D s / 3D Classic: 501 mm (19.7 in.)
3D Plus / 3D Mid: 574 mm (22.6 in.)
Ceph 1700 mm (66.9 in.)
15 Technical data for Planmeca ProMax product family
User's manual Planmeca ProMax 111