SUPERVISOR PROCEDURES
PROCLEIX
®
TIGRIS System Operator’s Manual 5-29
Volume I: Operating the Instrument
Part # 902649EN Rev. A
Line Content for HBV Discriminatory LIS Files
Each tube data line in the HBV Discriminatory LIS export contains the following fields,
separated by a comma or tab (literal values are given in quotes):
Field Contents Max Length
Sample ID For specimens, this is the barcode ID. Calibrators,
PROCLEIX
®
TIGRIS Controls, and External Quality
Controls (EQCs) are labeled as follows, where
“xxxxxx” is the six-digit lot number included in the
barcode ID:
♦ Negative Calibrator: “Negative Calibrator –
Lot xxxxxx”
♦ HBV Positive Calibrator: “HBV Positive
Calibrator – Lot xxxxxx”
♦ Negative EQC: “EQC Negative – Lot
xxxxxx”
♦ HBV Positive EQC: “EQC HBV Positive –
Lot xxxxxx”
♦
PROCLEIX ULTRIO
®
or PROCLEIX
ULTRIO
Plus
®
TIGRIS Negative Control:
“Negative Control – Lot xxxxxx”
♦
PROCLEIX ULTRIO
®
or PROCLEIX
ULTRIO
Plus
®
TIGRIS HBV Control: “HBV
Positive Control – Lot xxxxxx”
59
Overall Interpretation Calibrators and PROCLEIX TIGRIS Controls:
“Valid” or “Invalid”
Specimens and EQCs:
“Invalid”, “NonReactive”, “Reactive”, “Suspect”
254 (13 typical)
Assay Name PROCLEIX
®
HBV Discriminatory Assay:
“dHBV”
PROCLEIX
ULTRIO
Plus
®
HBV Discriminatory
Assay:
“UP dHBV”
The assay name in the LIS file may not match on-
screen references.
254 (15 typical)
Worklist ID Worklist ID (Run number). 6
Run Date & Time Date and time run was completed in the format:
YYYY-MM-DD HH:MM:SS
19
Tube Status Codes Status codes (if any) associated with the tube.
Codes are separated by a single space. See
"Appendix A. Status Codes" for code definitions.
254 (10 typical)
Internal Control RLU RLU value as a string. 20
Internal Control Result “Valid” or “Invalid”. 254 (7 typical)
Analyte RLU RLU value as a string. 20
Analyte S/CO S/CO as a string. 20
Kinetic Index Kinetic Index as a string. 20
Operator Username Login Username of the operator. 8
Internal Control Cutoff Minimum Internal Control RLUs for Valid IC. 20
Analyte Cutoff Minimum RLUs required from specimen to be
reactive.
20