SECTION 5
5-28 PROCLEIX
®
TIGRIS System Operator’s Manual
Part # 902649EN Rev. A Volume I: Operating the Instrument
Table 5.6 HCV Discriminatory LIS File Line Content
Example tube data line for a HCV Discriminatory LIS File (comma delimited):
MAS5632,Reactive,dHCV,100377,2005-01-06
22:08:09,IUO,52177,Valid,1372873,22.64,1.13814,miked,92618,60628,0.000,185237.333,,,1515
702.333,185237.667,311348,2025-12-31 23:59:59,HC+1027,HC+E5.11 12/02/04,10210 Genetic
Center Dr,Specimen,,,100377-002575-4-0,2.2.18,2.2.18
All of the information above appears on a single line in the PROCLEIX TIGRIS System LIS file.
Negative Calibrator
Analyte Average
Average RLU value for Negative Calibrator Analyte. 20
Negative Calibrator IC
Average
Average RLU value for Negative Calibrator IC. 20
Blank Reserved. 0
Blank Reserved. 0
HCV Positive Calibrator
Analyte Average
Average RLU value for HCV Positive Calibrator
Analyte.
20
HCV Positive Calibrator
IC Average
Average RLU value for Positive HCV Calibrator IC. 20
Master Lot Number Master Lot information as entered on the Inventory
Status screen.
6
Master Lot Expiration
Date
Master Lot Expiration Date information as entered in
Inventory Status
screen.
19
Instrument serial # Serial # of the instrument, with the prefix “HC+”. 254 (15 typical)
Firmware Revision Firmware version, with the prefix “HC+”. 254 (20 typical)
Site Information Identifying site information. 512
Type of tube (e.g.,
“Negative Calibrator”)
Negative Calibrator, HCV Positive Calibrator,
Specimen, Negative Control, HCV Positive Control.
254 (10 typical)
Blank Reserved. 0
Blank Reserved. 0
Test ID Test ID in the format: WorklistID-RackBarcode-
PositionInRack-ReplicateNum.
18
ADM Version Number ADM Version Number. 254 (10 typical)
TIGRIS Software
Version Number
PROCLEIX TIGRIS System Software Version
Number.
254 (10 typical)