RUNNING SAMPLES
PROCLEIX
®
TIGRIS System Operators Manual 3-7
Volume I: Operating the Instrument Part # 902649EN Rev. A
Table 3.2 shows an example of where calibrators, PROCLEIX TIGRIS Controls, and
specimens should be placed for a worklist with 202 specimens, using the default control
bracket size of 172 specimens. This example worklist requires 11 sample racks.
Table 3.2 Example Tube Placement
Figure 3.3 Sample Pipetting Order
Rack #
Rack
Position
Sample Tube Replicates Tube Barcode
1 1 Negative Calibrator 3 0425…
1 2 Positive Calibrator 3 0427…
1 3-20 Specimens 1-18 1 TBD
2 1-20 Specimens 19-38 1 TBD
3 1-20 Specimens 39-58 1 TBD
4 1-20 Specimens 59-78 1 TBD
5 1-20 Specimens 79-98 1 TBD
6 1-20 Specimens 99-118 1 TBD
7 1-20 Specimens 119-138 1 TBD
8 1-20 Specimens 139-158 1 TBD
9 1-14 Specimens 159-172 1 TBD
9 15 Negative Control 1 0406...
9 16 Positive Control 1 0410...
9 17-20 Specimens 173-176 1 TBD
10 1-20 Specimens 177-196 1 TBD
11 1-6 Specimens 197-202 1 TBD
11 7 Negative Control 1 0406…
11 8 Positive Control 1 0410…
Note: This example is for a single analyte assay run. Other assays may have different
calibrators, controls, replicates, and tube barcodes. See the appropriate PROCLEIX Assay
package insert for further instructions.