SECTION 1
1-22 PROCLEIX
®
TIGRIS System Operators Manual
Part # 902649EN Rev. A Volume I: Operating the Instrument
Water for the PROCLEIX
®
TIGRIS System Specifications
* Temperature compensated to 25° C
** CFU = Colony Forming Unit
Table 1.10 Water for the PROCLEIX
®
TIGRIS System Specifications
Power Requirements
Table 1.11 Power Requirements
Vacuum Pressure Requirements
The maximum vacuum pressure is 12 inches Hg.
Table 1.12 Minimum Vacuum Pressure
Sample Tube Specifications
Table 1.13 Sample Tube Requirements
Refer to the appropriate PROCLEIX
®
Assay package insert for more information about
specimen collection tubes.
Electrical resistivity* >
10 ΩM-cm
Bacteria <
10 CFU/mL**
Particulate matter 0.22 micron filter
Total Oxidizable Organic Carbon < 0.05 mg/L
Electrical Input 200 – 240 VAC, 50-60 Hz
Input Current 16 A
Electrical Connector L6-30R (USA)
Local Altitude Minimum Pressure
1 to 2000 feet (0 to 610 m) 6.5 inches Hg
2001 to 4000 feet (611 to 1219 m) 6.0 inches Hg
4001 to 5000 feet (1220 to 1524 m) 5.5 inches Hg
5001 to 7355 feet (1525 to 2242 m) 5.0 inches Hg
Type Size Material
Specimen
12 x 75mm to 16 x 100mm (including
13 x 75mm, 13 x 100mm, and 16 x
75mm, 13 x 82 mm)
Polypropylene plastic or
siliconized glass
Calibrators,
PROCLEIX
®
TIGRIS
Controls
13mm x 82mm Polypropylene plastic