SUPERVISOR PROCEDURES
PROCLEIX
®
TIGRIS System Operator’s Manual 5-5
Volume I: Operating the Instrument
Part # 902649EN Rev. A
Software Configuration
System configuration parameters are grouped together in the Options box on the Configuration
tab.
To enable an option, check the box next to it. The system automatically saves the changes.
Table 5.2 describes available options.
*If this option is not enabled, the instrument will not allow “Flag Duplicate Samples” to be enabled.
Table 5.2 Configuration Options
Option Description
Required
Setting
Auto Logoff after no
activity for:__ minutes
Automatically logs off the current operator after the
specified amount of time without activity.
No
requirement
Require Double Entry of
Sample IDs for Test
Requests
Requires double entry for manual keyboard entry. Enabled
Scanning Barcode
Confirms Sample ID
Not applicable to PROCLEIX Assays. Disabled
Fix Unreadable Barcodes
before Processing*
Allows the operator to fix unreadable barcodes before
processing. Prevents the system from automatically
assigning a Sample ID; all barcodes must be
successfully scanned or manually entered. If this
option is not enabled, the system automatically
assigns a Sample ID in the case of an unreadable
barcode.
Enabled
Auto Start Pipetting after
Rack Validation
The system automatically starts pipetting after racks
have been validated.
Enabled
Display Tip Count Displays the pipette tip count on the
Inventory screen. Enabled
Place Controls every __
Samples
Sets the number of specimens between PROCLEIX
®
TIGRIS Controls when control bracketing is being
used. See “Control Bracketing” on page 3-2 for more
information. The default setting is 172 specimens, but
this can be changed to any value between 1 and 492
(single analyte assay) or 1 and 487 (triple analyte
assay).
No
requirement
Require Operator Logon
When Worklist Queued
Requires the operator to re-enter a user ID and
password to queue a worklist.
No
requirement
Flag Duplicate Samples Flags sample barcodes that have been run on a
previous or in-process worklist.
Enabled
Enable Manually Created
Worklists
Not applicable to PROCLEIX Assays. Disabled