SECTION 2
2-4 PROCLEIX
®
TIGRIS System Operator’s Manual
Part # 902649EN Rev. A Volume I: Operating the Instrument
Status Bar
The Status Bar displays the system mode, current user, current system action, and the date/
time. Use the system mode to determine the availability of the system for certain activities.
Table 2.4 describes possible system modes.
Table 2.4 System Modes
Mode Meaning Occurrence
Assay Processing
The system is processing a
worklist.
The system is in this mode as worklists
are queued for processing. This mode
continues through specimen pipetting,
assay processing, and results
processing.
Data Access Only
The system is busy, but some
software functions are
available.
The system is in this state during system
initialization, hardware reset, and during
unload and priming operations.
Data Initialization
The system and software are
initializing. The system is in
this state at start-up.
A progress bar is displayed during this
state and no program features or
functions are available.
Error—Data
Access Only
A system error has occurred.
The operator can manipulate
worklists and data in the
system.
The system is in this state if an error
occurs while the system is connecting to
the system, performing a hardware reset
or performing an unload operation.
Error—No Data
Access
A system error has occurred.
The operator cannot perform
any data access activities
except Send Logs.
The system is in this state if an error
occurs for any other reason than the
system connecting to the system,
performing a hardware reset or
performing an unload operation.
Standby
The system is ready for
normal operation.
The system is in this state after the
system and software have been
initialized or have returned to an idle
state.