APPENDIX A
A-4 PROCLEIX
®
TIGRIS System Operator’s Manual
Part # 902649EN Rev. A Volume I: Operating the Instrument
Results Processing Errors
TLH A Left Heater temperature was out of
range.
This is a non-fatal error. All samples in
the affected MTU are marked invalid.
TRH A Right Heater temperature was out of
range.
This is a non-fatal error. All samples in
the affected MTU are marked invalid.
TS The ambient temperature in the Sample
Carousel area was out of range.
Sample pipetting pauses, but assay
processing continues. No MTUs are
marked Invalid. Sample pipetting
continues when the gauge returns to the
ambient temperature.
The affected samples are marked Invalid.
URM MTU barcode error. This is a non-fatal error. All samples in
the affected MTU are marked Invalid.
This error does not count towards the
10% invalid criterion.
VVFS The volume of the wTCR/Sample in the
MTU tube was higher or lower than
expected.
This is a non-fatal error. The affected
samples are marked Invalid.
VVFT The volume of the wTCR in the MTU tube
was higher or lower than expected.
This is a non-fatal error. The affected
samples are marked Invalid.
Note: The code descriptions in this table may not match those that appear in the legend
on the last page of the Results report. The descriptions listed here provide greater detail
regarding the specific cause of individual errors.
Code Description Effect
af Post-kinetic index adjustment flasher
out of range for calibrator or control.
Test invalid.
afc Post-kinetic index adjustment specimen
flasher greater than cutoff.
Test invalid.
ag Post-kinetic index adjustment glower
out of range for calibrator or control.
Test invalid.
agc Post-kinetic index adjustment specimen
glower greater than cutoff.
Test invalid.
cv % cv out of range. Test invalid.
e Data analysis error. Test invalid.
f Failed run flag. Test invalid.
fa Flasher average out of range. Test invalid.
fb Front to back ratio out of range. Test invalid.
fl Calibrator flasher out of range. Test invalid.
g Calibrator glower out of range. Test invalid.
ga Glower average out of range. Test invalid.
ic Invalid sample became valid due to a
crossover.
Test invalid.
Code Cause Effect