STATUS CODES
PROCLEIX
®
TIGRIS System Operator’s Manual A-5
Volume I: Operating the Instrument
Part # 902649EN Rev. A
m Early maximum allowable RLU. Test invalid.
p Assay Processing Error. Test invalid.
pn Negative calibrator peak out of range. Test invalid.
pp Positive calibrator peak out of range. Test invalid.
r R-squared out of range. Test invalid.
sl Slope out of range. Test invalid.
sp Spike detected. Test invalid.
t RLU total out of range for calibrators or
specimens.
Test invalid.
ta Total average out of range. Test invalid.
ts Invalid calculation occurred. Test invalid.
v Too many IC failures in this bracket. Valid “Nonreactive” samples within the
bracket are marked “Suspect”.
x Invalidated or flagged as “Suspect” by
failed PROCLEIX
®
TIGRIS Controls.
Valid “Nonreactive” samples within the
bracket are marked “Suspect”.
Code Description Effect