SECTION 5
5-32 PROCLEIX
®
TIGRIS System Operators Manual
Part # 902649EN Rev. A Volume I: Operating the Instrument
Table 5.8 PROCLEIX
®
WNV Assay LIS Files
Example tube data line of PROCLEIX WNV Assay LIS File (comma delimited):
MAS5698,NonReactive,WNV,100002,2004-03-30
02:04:28,IUO,235623,Valid,0,0.00,1.33942,miked,90024,44947,1063.000,180048.333,1462805.0
00,180317.667,,,032504,2025-12-31 23:59:59,HC+1030,HC+E5.01 02/20/04,10210 Genetic
Center Dr,Specimen,,,100002-001357-12-1,1.6.0.10.001,1.6.0.9
All of the above information appears on a single line in the file.
Configuring the Archive Parameters
Use the Archive Configuration to set archive parameters, including the archive path and auto-
archive frequency. An operator must have “Maintenance - Software Setup” rights to configure
the archive parameters.
The following data is saved to the location specified by the archive path, which may be the
PROCLEIX
®
TIGRIS System computer hard drive, a CD-RW, or a USB flash memory drive:
all log files
kinetic data
results reports
worklists
database archive
LIS files are not saved to this location. For information on LIS data location, see “Configuring
the LIS” on page 5-19.
Positive Calibrator IC
Average
Average RLU value for Positive Calibrator IC. 20
Blank Reserved. 0
Blank Reserved. 0
Master Lot Number Master Lot information as entered on the Inventory
Status screen.
6
Master Lot Expiration
Date
Master Lot Expiration Date information as entered in
Inventory Status
screen.
19
Instrument serial # Serial # of the instrument, with the prefix “HC+”. 254 (15
typical)
Firmware Revision Firmware version, with the prefix “HC+”. 254 (20
typical)
Site Information Identifying site information. 512
Type of tube (e.g.,
“Negative Calibrator”)
Negative Calibrator, Positive Calibrator, Specimen,
Negative Control, Positive Control
254 (10
typical)
Blank Reserved. 0
Blank Reserved. 0
Test ID Test ID in the format: WorklistID-RackBarcode-
PositionInRack-ReplicateNum.
18
ADM Version Number ADM Version Number. 254 (10
typical)
TIGRIS Software
Version Number
PROCLEIX TIGRIS System Software Version
Number.
254 (10
typical)