SUPERVISOR PROCEDURES
PROCLEIX
®
TIGRIS System Operator’s Manual 5-31
Volume I: Operating the Instrument
Part # 902649EN Rev. A
Line Content for the PROCLEIX
®
WNV Assay LIS Files
Each tube data line in the PROCLEIX WNV Assay LIS export contains the following fields,
separated by a comma or tab (literal values are given in quotes)
Field Contents Max Length
Sample ID For specimens, this is the barcode ID. Calibrators,
PROCLEIX
®
TIGRIS Controls, and External Quality
Controls (EQCs) are labeled as follows, where
xxxxxx” is the six-digit lot number included in the
barcode ID:
Negative Calibrator: “Negative Calibrator –
Lot xxxxxx
Positive Calibrator: “Positive Calibrator –
Lot xxxxxx
Negative EQC: “EQC Negative – Lot
xxxxxx
Positive EQC: “EQC Positive – Lot xxxxxx
PROCLEIX
®
WNV TIGRIS Negative
Control: “Negative Control – Lot xxxxxx
PROCLEIX
®
WNV TIGRIS Positive
Control: “Positive Control – Lot xxxxxx
59
Overall Interpretation Calibrators and PROCLEIX TIGRIS Controls:
“Valid” or “Invalid”
Specimens and EQCs:
“Invalid”, “NonReactive”, “Reactive”, “Suspect”
254 (13
typical)
Assay Name WNV”
The assay name in the LIS file may not match on-
screen references.
254 (15
typical)
Worklist ID Worklist ID (Run number). 6
Run Date & Time Date and time run was completed in the format:
YYYY-MM-DD HH:MM:SS
19
Tube Status Codes Status codes (if any) associated with the tube. Codes
are separated by a single space. See "Appendix A.
Status Codes" for code definitions.
254 (10
typical)
Internal Control RLU RLU value as a string. 20
Internal Control Result “Valid” or “Invalid”. 254 (7 typical)
Analyte RLU RLU value as a string. 20
Analyte S/CO S/CO as a string. 20
Kinetic Index Kinetic Index as a string. 20
Operator Username Login Username of the operator. 8
Internal Control Cutoff Minimum Internal Control RLUs for Valid IC. 20
Analyte Cutoff Minimum RLUs required from specimen to be
reactive.
20
Negative Calibrator
Analyte Average
Average RLU value for Negative Calibrator Analyte. 20
Negative Calibrator IC
Average
Average RLU value for Negative Calibrator IC. 20
Positive Calibrator
Analyte Average
Average RLU value for Positive Calibrator Analyte. 20