RUNNING SAMPLES
PROCLEIX
®
TIGRIS System Operators Manual 3-3
Volume I: Operating the Instrument Part # 902649EN Rev. A
appear on the Results screen as each bracket finishes processing. The software can also be
configured to automatically release results to the LIS as each bracket finishes processing. See
“Configuring System Settings” on page 5-4 for more information about configuration. See the
appropriate PROCLEIX
®
Assay package insert for more information about results interpretation
based on control bracketing.
Creating a Worklist
To create a Carousel Worklist:
1. From either the Worklist Manager
or Run Manager screen, click Create Carousel
Worklist.
A Create Carousel Worklist
window opens.
2. Select the assay to be run from the Assay drop-down list.
The Analyte drop-down list defaults to the appropriate analyte.
3. In the Total Specimens text box, enter the total number of specimens that are to be
included in the worklist. Include custom user-defined controls in the total specimen count,
but do not include the required PROCLEIX
®
TIGRIS Controls or calibrators.
4. In the Replicates box, type the number of replicates for each sample to be tested.
The Total Tests in the upper right hand corner of the Create Carousel Worklist
window
reflects the total specimens plus required calibrator and PROCLEIX TIGRIS Control
replicates.
5. Click OK.
The system adds the worklist to the Stored Worklists
grid.
A Carousel Worklist does not have any Sample IDs associated with it until the racks are
loaded and scanned.
Editing a Worklist
If a worklist has not yet been queued and scanned, the Worklist Manager screen can be used
to edit it:
1. Select the worklist to be edited from the Stored Worklists
grid.
2. Click Edit Carousel Worklist....
An Edit Carousel Worklist
window opens.
3. Use the Edit Specimen Count... button to change the number of specimens in the
worklist:
a) Click Edit Specimen Count....
An Edit Specimen Count
window opens.
b) In the New Specimen Count text box, type the new number of specimens for the
worklist.
c) Click OK.
Note: When the control bracket size is changed in the software configuration, the new
bracket size does not apply retroactively to existing worklists.
Note: The Edit Carousel Worklist button is grayed out until a Carousel Worklist is
selected.