2-11
VIBXPERT 11.2010
Getting started - Memory card
When the device is switched on, the card is automatically tested. The
following scenarios are possible:
Card formatted, which version?
- The card is new and has not yet been formatted in a VIBXPERT:
you are prompted to format the card.
- The card has already been used in another VIBXPERT:
Card version < firmware version in the device: the files on the
card are automatically updated.
Card version > firmware version in the device: you are prompted
to reset all settings to the "factory settings" ('Reset', P. 2-29).
Card version = latest firmware version: no action.
Card faulty?
- The file system on the card is damaged. A warning appears along
with the instruction to have the card repaired. If possible, make a
backup copy of the measurement data (see 'Backup', page 6-9).
Start card formatting and repair in the 'Service' menu (P. 2-29).
How much free memory capacity is there?
The free memory capacity is checked regularly during operation. A
message appears when the memory is almost full.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the PRÜFTECHNIK VIB 5.300 Series and is the answer not in the manual?

PRÜFTECHNIK VIB 5.300 Series Specifications

General IconGeneral
BrandPRÜFTECHNIK
ModelVIB 5.300 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals