4-3
VIBXPERT 11.2010
Results
RMS
0-p
p-p
Zoomed area (gray)
Characteristic overall values - trending
If more than two measurements are saved in a file, the results for
each overall value are displayed as a trend curve (see p. 3-28f).
Markers on the trend curve indicate each individual measurement.
The measurement values at the cursor position, the date, the rpm if
applicable, and assigned events and comments are specified below
the diagram. Use the F key to display the individual results.
With the aid of the displayed threshold line (see p. 3-25), it is
possible to determine if thresholds have been exceeded. If more than
one overall value appears in the diagram, only the threshold for the
highlighted value is displayed (in the below screen: upper alarm for
p-p value). Move the joystick up / down to mark the other overall
values and display the respective threshold values.
Zoom
Press the '+' button repeatedly to magnify the area around the cursor.
This process can be undone again with the 'Minus(–)' key.
The gray bars above a diagram show the displayed range relative to
the entire display range.
Deleting individual measurements
If a spurious measurement has been included in the trend, delete this
as follows:
Highlight the measurement in the trend with the cursor.
Press the MENU button.
Click on 'Delete'.
F
Left:
Delete measurement
Right:
Trend diagram

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the PRÜFTECHNIK VIB 5.300 Series and is the answer not in the manual?

PRÜFTECHNIK VIB 5.300 Series Specifications

General IconGeneral
BrandPRÜFTECHNIK
ModelVIB 5.300 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals