494 Appendix A Command Reference
PSPG
Probe Sync Pulse Gate
Class Special Sync Probe Pulse
Description Enables and disables the probe pulse on the special sync output. The PSPG? query
returns the current PSPG mode.
Command syntax PSPD mode
mode integer
0 = OFF
1 = ON
Query syntax PSPG?
Query response 0 or 1
Other required
commands
The FMTU command instructs the generator to use the new settings. The ALLU command
updates the hardware to the new settings and redraws the test image.
Related commands PSPP, PSPW, PSHD, PSVW, PSPM
Example
FSPG 0 // Disables frame sync
PSPG 1 // Enables probe pulse on special sync
ALLU // Applies the settings