EasyManuals Logo

Rigol DS1104Z Programming Guide

Rigol DS1104Z
172 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #145 background imageLoading...
Page #145 background image
Chapter 2 Command System RIGOL
DS1000Z Programming Guide 2-131
:TRIGger:SPI:TIMeout
Syntax :TRIGger:SPI:TIMeout <time_value>
:TRIGger:SPI:TIMeout?
Description Set or query the timeout value when the trigger condition is TIMeout in SPI trigger. The
default unit is s.
Parameter
Name
Type
Range
Default
<time_value>
Real
100ns to 1s
1μs
Return
Format
The query returns the timeout value in scientific notation.
Example
:TRIGger:SPI:TIMeout 0.001 /*Set the timeout value to 1ms*/
:TRIGger:SPI:TIMeout? /*The query returns 1.000000e-03*/
Related
Command
:TRIGger:SPI:WHEN
:TRIGger:SPI:SLOPe
Syntax :TRIGger:SPI:SLOPe <slope>
:TRIGger:SPI:SLOPe?
Description
Set or query the clock edge in SPI trigger.
Parameter
Name
Type
Range
Default
<slope> Discrete {POSitive|NEGative} POSitive
Explanation POSitive: sample the SDA data on the rising edge of the clock.
NEGative: sample the SDA data on the falling edge of the clock.
Return
Format
The query returns POS or NEG.
Example
:TRIGger:SPI:SLOPe POSitive /*Set the clock edge to POSitive*/
:TRIGger:SPI:CLEVel
Syntax :TRIGger:SPI:CLEVel <level>
:TRIGger:SPI:CLEVel?
Description Set or query the trigger level of the SCL channel in SPI trigger. The unit is the same as the
current amplitude unit.
Parameter
Name
Type
Range
Default
<level>
Real
(± 5 × VerticalScale from the screen center) - OFFSet
0
Explanation For VerticalScale, refer to :CHANnel<n>:SCALe. For OFFSet, refer
to :CHANnel<n>:OFFSet.
Return
Format
The query returns the trigger level in scientific notation.
Example
:TRIGger:SPI:CLEVel 0.16 /*Set the trigger level to 160mV*/
:TRIGger:SPI:CLEVel? /*The query returns 1.600000e-01*/
Related
Commands
:CHANnel<n>:SCALe
:CHANnel<n>:OFFSet

Table of Contents

Other manuals for Rigol DS1104Z

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS1104Z and is the answer not in the manual?

Rigol DS1104Z Specifications

General IconGeneral
Bandwidth100 MHz
Channels4
Sample Rate1 GSa/s
Waveform Capture Rate30, 000 wfms/s
Vertical Sensitivity1 mV/div to 10 V/div
Time Base Range5 ns/div to 50 s/div
Vertical Resolution8 bits
InterfaceUSB Host, USB Device, LAN, AUX
Display7 inch TFT LCD (800 x 480)
Trigger TypesEdge, Pulse, Video, Slope
Power Supply100-240 V, 45-440 Hz
Memory Depth24 Mpts (Single Channel), 12 Mpts (Dual Channel)

Related product manuals