EasyManuals Logo

Rigol DS1104Z Programming Guide

Rigol DS1104Z
172 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #17 background imageLoading...
Page #17 background image
Chapter 2 Command System RIGOL
DS1000Z Programming Guide 2-3
:SINGle
Syntax
:SINGle
Description Set the oscilloscope to the single trigger mode. This command is equivalent to the
following two operations: pressing the SINGLE key at the front panel and sending
the :TRIGger:SWEep SINGle command.
Explanation In the single trigger mode, the oscilloscope triggers once when the trigger conditions
are met and then stops.
When the waveform record function is enabled or during the playback of the
recorded waveform, this command is invalid.
Related
Commands
:TFORce
:RUN
:STOP
:TFORce
Syntax
:TFORce
Description Generate a trigger signal forcefully. This command is only applicable to the normal and
single trigger modes (see the
:TRIGger:SWEep command) and is equivalent to pressing
the FORCE key at the front panel.
Related
Command
:TRIGger:SWEep

Table of Contents

Other manuals for Rigol DS1104Z

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS1104Z and is the answer not in the manual?

Rigol DS1104Z Specifications

General IconGeneral
Bandwidth100 MHz
Channels4
Sample Rate1 GSa/s
Waveform Capture Rate30, 000 wfms/s
Vertical Sensitivity1 mV/div to 10 V/div
Time Base Range5 ns/div to 50 s/div
Vertical Resolution8 bits
InterfaceUSB Host, USB Device, LAN, AUX
Display7 inch TFT LCD (800 x 480)
Trigger TypesEdge, Pulse, Video, Slope
Power Supply100-240 V, 45-440 Hz
Memory Depth24 Mpts (Single Channel), 12 Mpts (Dual Channel)