EasyManuals Logo
Home>Rigol>Test Equipment>DS1104Z

Rigol DS1104Z Programming Guide

Rigol DS1104Z
172 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #22 background imageLoading...
Page #22 background image
RIGOL Chapter 2 Command System
2-8 DS1000Z Programming Guide
:CALibrate Commands
Command List:
:CALibrate:QUIT
:CALibrate:STARt
:CALibrate:QUIT
Syntax
:CALibrate:QUIT
Description
Exit the calibration at any time.
Related
Command
:CALibrate:STARt
:CALibrate:STARt
Syntax
:CALibrate:STARt
Description The oscilloscope starts to execute the self-calibration.
Explanation The self-calibration can make the oscilloscope quickly reach its optimum working
state to obtain the most accurate measurement values.
During the self-calibration, all the channels of the oscilloscope must be disconnected
from the inputs.
The functions of most of the keys are disabled during the self-calibration. You can
send the :CALibrate:QUIT command to quit the self-calibration.
Related
Command
:CALibrate:QUIT

Table of Contents

Other manuals for Rigol DS1104Z

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS1104Z and is the answer not in the manual?

Rigol DS1104Z Specifications

General IconGeneral
BrandRigol
ModelDS1104Z
CategoryTest Equipment
LanguageEnglish

Related product manuals