EasyManuals Logo
Home>Rigol>Test Equipment>DS1104Z

Rigol DS1104Z Programming Guide

Rigol DS1104Z
172 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #90 background imageLoading...
Page #90 background image
RIGOL Chapter 2 Command System
2-76 DS1000Z Programming Guide
Explanation These commands are used to select the waveform shape.
NOISe: select Noise
PULSe: select Pulse
RAMP: select Ramp
SINusoid: select Sine
SQUare: select Square
USER: select Arbitrary waveform
<freq>: set the frequency of the specified waveform (Noise does not have this
parameter). The default unit is Hz
<amp>: set the amplitude of the specified waveform. The default unit is Vpp
<offset>: set the DC offset of the specified waveform. The default unit is V
<phase>: set the start phase of the specified waveform (Noise does not have this
parameter). The default unit is degree (°).
Users can omit one or multiple parameters. When all of the parameters are omitted,
the commands only select the waveform shape but do not modify the parameters.
The four parameters (<freq>, <amp>, <offset> and <phase>) are in order. You can
not omit one parameter and set the one/ones behind it. For example, you can not
omit <freq> and set <amp> directly.

Table of Contents

Other manuals for Rigol DS1104Z

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS1104Z and is the answer not in the manual?

Rigol DS1104Z Specifications

General IconGeneral
BrandRigol
ModelDS1104Z
CategoryTest Equipment
LanguageEnglish

Related product manuals