Cedex Bio analyzer 4Software
Key screens
The color of a test button represents its status:
The test is on board and ready for use.
The test is blocked for one of the following reasons:
o The calibration is due or failed.
o For the reagent set, the number of available tests is 0, or a reagent bottle is missing
(incomplete reagent set).
o Initial calibration is required.
There are only few tests left.
A QC is due or its result has not been accepted.
A more recent version of the application has been imported.
The test is not on board.
A required diluent or cleaner is not on board.
A Test selection screen in order mode Easy. B Test selection screen in order mode Full.
Figure A-54 Test selection screen
A
B