Roche Diagnostics
B-80
Operatorโ€™s Manual ยท Version 3.4
5 Daily operation Cedex Bio analyzer
Performing QC
3
Press to look at result details.
4
Press to close the screen.
5
Press .
6
Do one of the following:
o Press Accept to accept the QC results for the selected test.
o Press Ignore to exclude the result from further QC result calculations such as
QC history statistics.
When ignoring a result, the due status does not change.
In the QC results tables, ignored QC results are marked with "I".
Flags of ignored results are not inherited to dependent results.
o Press Retransmit to send the result again.
(This option is active if your analyzer is connected to a host system.)
A Test name
B Accepted by: User name. $SYS$ means
automatically accepted.
C Control name
D Calculation date and time.
E Result
F Time when accepted.
G Press to display context information of the
fluids used to perform QC for this test.
Figure B-26
G
A
C
F
B
E
D
If the QC result is outside the defined range, perform the QC again. If the results are
still outside the range, check for other causes. If all fails, perform a calibration.