Roche Diagnostics
Operatorโs Manual ยท Version 3.4 B-171
Cedex Bio analyzer 7 Configuration
Configuration
Function User level
0 1 2 3 4
Button, screen
Not logged in
Operator
Supervisor
Lab Manager
Administrator
Disk Inventory Reagents can be put on the instrument
displayed and printed.
xllll
ISE Status ISE status can be displayed and
printed.
lllll
Cuvettes Cuvettes can be added lllll
System Status
regarding bottles
Can refill bottles. lllll
STAT Create and start a priority order lllll
Te st s Display, print test status, display
related information
lllll
Prepare xllll
Prepare Access the following screens:
o Check Bottles
o ISE Status
o Check Maintenance
o Disk Inventory
o Cuvettes
o Select Mixing
o Select Cal
o Cal Placement
xllll
Order
(in Overview screen)
xllll
End Shift xllll
Table B-22 User levels and access rights, Overview tab
Function User level
0 1 2 3 4
Button, screen
Not logged in
Operator
Supervisor
Lab Manager
Administrator
All screens in
Workplace tab
Display, print lllll
Orders screen Display, print, filter function lllll
Orders Create, delete xllll
QC Status Accept, ignore lllll
QC Status Perform single interval QC
measurement
xxlll
Table B-23 User levels and access rights, Workplace tab