Cedex Bio analyzer 10 Messages and alarms
List of alarm messages
7108 [%s] A not specified hardware error (%d) was reported by
the DC-Slave control.
Restart the system. If the error recurs export the log files
with the system remaining in this error condition. If the
error persists call service.
7109 [%d] The sample probe is bent or not properly aligned. Check the probe and replace it if necessary. Restart the
system. If the error persists call service.
7110 [%d] A timing error occurred while fluid is expelled into a
cuvette. Probably the error was caused by a previous
hardware problem.
Check for previous errors. Restart the system. If the error
persists call service.
7111 [%d] Transfer movement failed due to open cover. Keep covers closed while the system is in Operating status.
7112 [%s] The ISE reference sensor detected air when ISE Ref
solution was transported.
Check the ISE Ref bottle filling level. Perform <Prime ISE
Reference and Calib.> and check the fluid flow. If the
error persists call service.
7113 [%d] A prime cycle failed due to an open cover. Ensure that all covers are closed while the system is being
initialized, in the Operating or Maintenance status.
7114 [%d] An initialization error occurred. The X-transfer zero
positions are invalid.
Check the X-transfer for obstructions. Ensure that all cov-
ers are closed. Restart the system. If the error persists call
service.
7115 [%d] Dispensing into the ISE tower stopped because the
ISE tower was not empty.
Check the ISE tower waste tubing. Check the ISE tower
outlet for obstructions. Check the pinch valves. If the
error persists call service.
7116 [%d] The needle crashed while the system was trying to
dispense fluid.
Switch off the analyzer. Check if the transfer or probe
movement is obstructed. Restart the system. If the error
persists call service.
7117 [%d] Consequential error caused by a previous hardware
or software problem.
Check previous errors. Restart the system. If the error
recurs export the log files with the system remaining in
this error condition. If the error persists call service.
7118 [%s] Rotor initialization error. The light barrier did not
react.
Switch off the analyzer. Remove the rotor and clean the
light barrier. Reinstall the rotor. Restart the system. If the
error persists call service.
7119 [%d] Consequential error caused by a previous hardware
or software problem.
Check previous errors. Restart the system. If the error
recurs export the log files with the system remaining in
this error condition. If the error persists call service.
7120 [%s] Either a rotor, an X-transfer or a Z-transfer position
error occurred. Maybe one of these devices has been
moved manually by accident.
Restart the system. If the error persists call service.
7121 [%d] Consequential error caused by an internal
communication problem.
Check previous errors. Restart the system. If the error
recurs export the log files with the system remaining in
this error condition. If the error persists call service.
7122 [%d] The Prepare action has been stopped due to an open
cover.
Ensure that all covers are closed while the system is being
initialized, in the Operating or Maintenance status.
7123 Extra wash cycle after reagent pipetting failed. Perform <Deproteinize Probe> before next run.
7124 Analyzer temperature outside the range.
7125 [%s] Motion control error in rotor motor. Restart the system to correct the problem. If the error per-
sists call service.
7126 [%s] Motion control error X-transfer motor. Restart the system to correct the problem. If the error per-
sists call service.
7127 [%s] Motion control error Z-transfer motor. Restart the system. If the error persists call service.
ID Message Comment
Table E-1
List of alarm messages (Continued)