10 Messages and alarms Cedex Bio analyzer
List of alarm messages
7128 [%s] Initialization error X-transfer. Switch off the analyzer. Ensure that all covers are closed.
Check if the X-transfer is blocked or jamming. Restart the
system. If the error persists call service.
7129 [%s] Initialization error Z-transfer. Switch off the analyzer. Check if the Z-transfer is blocked
or jamming. Check the seating of the transfer head cover.
Restart the system. If the error persists call service.
7130 [%s] Rotor position error. Switch off the analyzer and check if the rotor is blocked or
jamming. If the error recurs call service.
7131 [%s] Position error X-transfer. Switch off the analyzer and check if the X-transfer is
blocked or jamming. If the error recurs call service.
7132 [%s] Initialization error Z-transfer. Switch off the analyzer. Check if the Z-transfer is blocked
or jamming. Check the seating of the transfer head cover.
Restart the system. If the error persists call service.
7133 [%s] Rotor motor overheated. Switch off the analyzer. Check if the rotor is blocked or
jamming. Restart the system. If the error persists call ser-
vice.
7134 [%s] Motor X-transfer overheated. Switch off the analyzer. Check if the X-transfer is blocked
or jamming. Ensure that the X-transfer can be moved
smoothly. Restart the system. If the error persists call ser-
vice.
7135 [%s] Motor Z-transfer overheated. Switch off the analyzer. Check if the Z-transfer is blocked
or jamming. Ensure that the Z-transfer can be moved
smoothly. Restart the system. If the error persists call ser-
vice.
7136 [%s] Tube bottom detector is blocked or jamming. Switch off the analyzer. Remove the transfer head cover.
Check the tube bottom detector mechanism. Install the
transfer head cover. Restart the system. If the error per-
sists call service.
7137 [%s] Na sample measurement unstable. Check the expiry date of the Ref and Na electrodes.
Replace expired electrodes. Check the ISE tubing for
obstructions and leaks. Perform <Electrode Service>. If
the error persists call service.
7139 [%s] K sample measurement unstable. Check the expiry date of the Ref and K electrodes. Replace
expired electrodes. Check the ISE tubing for obstructions
and leaks. Perform <Activate Electrodes>. If the error per-
sists call service.
7140 [%s] Na calibrator measurement unstable. Perform <Electrode Service> and <Prime ISE Reference
and Calib.>. Check ISE Cal and Ref flow. Check the ISE
tubing for obstructions. If the error persists call service.
7142 [%s] K calibrator measurement unstable. Perform <Electrode Service> and <Prime ISE Reference
and Calib.>. Check ISE Cal and Ref flow. Check the ISE
tubing for obstructions. If the error persists call service.
7143 [%d] An internal software problem (%d) occurred. The
access to the internal EEPROM failed.
Restart the system. If the error recurs export the log files
with the system remaining in this error condition. If the
error persists call service.
7144 [%d] The instrument battery is low or not connected. The
date and time may no longer be valid. A replacement of
the battery is probably required.
Call service.
7145 [%d] The action <Clean ISE> has been stopped due to an
open cover.
Ensure that all covers are closed while the system is being
initialized, or in the Operating or Maintenance status.
ID Message Comment
Table E-1
List of alarm messages (Continued)