Roche Diagnostics
Operatorโs Manual ยท Version 3.4 E-19
Cedex Bio analyzer 10 Messages and alarms
List of alarm messages
7146 [%d] The initialization of the system has been stopped due
to an open cover.
Ensure that all covers are closed while the system is being
initialized, or in the Operating or Maintenance status.
7147 [%d] The initialization of the ISE Unit has been stopped
due to an open cover.
Ensure that all covers are closed while the system is being
initialized, or in the Operating or Maintenance status.
7148 [%d] Initialization of the ISE Unit has been stopped due to
a blocked transfer.
Switch off the analyzer. Check if the transfer is blocked or
jamming. Ensure that all covers are closed. Restart the sys-
tem. If the error persists call service.
7149 [%d] Initialization of the rotor failed. The calculated offset
is outside the range.
Switch off the analyzer. Check if the rotor is blocked or
jamming. Restart the system. If the error persists call ser-
vice.
7150 [%d] Initialization of the transfer failed due a level
detection problem.
Switch off the system. Clean the probe and probe holder
and ensure that they are installed correctly. Restart the
system. If the error persists call service.
7151 [%d] Dispensing fluid into the ISE tower failed due to an
incorrect X-transfer or probe position.
Switch off the analyzer. Check if the X- transfer and Z-
transfer are blocked or jamming. Restart the system. If the
error persists call service.
7152 [%d] A pipetting or dispensing cycle failed due to an
incorrect X-transfer or probe position.
Switch off the analyzer. Check if the X- transfer and Z-
transfer are blocked or jamming. Restart the system. If the
error persists call service.
7153 [%d] A reagent mixing cycle failed because the probe
crashed.
Switch off the analyzer. Check if the probe and probe
holder are correctly installed. Replace a bent probe.
Restart the system. If the error persists call service.
7154 [%d] A probe wash cycle failed because the probe crashed. Switch off the analyzer. Check if the probe and probe
holder are correctly installed. Replace a bent probe.
Restart the system. If the error persists call service.
7155 [%d] Initialization of the transfer failed because the probe
crashed.
Switch off the analyzer. Check if the probe and probe
holder are correctly installed. Replace a bent probe.
Restart the system. If the error persists call service.
7156 [%d] A prime cycle failed because the probe crashed. Switch off the analyzer. Check if the probe and probe
holder are correctly installed. Replace a bent probe.
Restart the system. If the error persists call service.
7157 [%d] Initialization of the Z-transfer failed due to a level
detection problem.
Switch off the analyzer. Check if the probe and probe
holder are correctly installed. Clean the init plate. Restart
the system. If the error persists call service.
7158 [%d] A tube bottom sensor problem was detected. Switch off the analyzer. Check if the tube bottom sensor
mechanism can be moved freely. Restart the system. If the
error persists call service.
7159 [%d] Movement stopped because the probe crashed. Switch off the analyzer. Check if the probe and probe
holder are correctly installed. Replace a bent probe. Check
the magnetic slider. Restart the system. If the error per-
sists call service.
7160 Sample pipetting on position %d was aborted due to an
unstable LD signal%s.
7161 Reagent pipetting of bottle set %s was aborted due to an
unstable level detection signal%s. The test is automatically
repeated.
7162 The probe crashed due to a jamming main cover slider. Check the slider and ensure that it is operating correctly.
7163 The run aborted due to a jamming main cover slider. Check the slider and ensure that it is operating correctly.