Cedex Bio analyzer 14 ISE operation
Replacing ISE fluid bottles
18
On the ISE status screen, press .
A placement list is displayed, telling you where to place ISE Solution 1 and 2.
19
Place ISE Solution 1 and ISE Solution 2 on the positions indicated on the screen.
20
Press .
The Overview tab is displayed.
21
Press to perform the main calibration.
22
When the system is in Standby status, remove ISE Solution 1, and ISE Solution 2
from the sample area.
23
Press .