172User Manual 1173.9463.02 ─ 06
Bit No. Meaning
6 User Request
This bit is set on pressing the LOCAL key, i. e. when the instrument is switched over to manual
control.
7 Power On (supply voltage on)
This bit is set when the instrument is switched on.
Related common commands
The Event Status Register (ESR) can be queried using ESR? / XESR?.
The Event Status Enable (ESE) register can be set using the command *ESE / *XESE
and read using *ESE? / *XESE?.
See also Common Commands
5.6.3.4 STATus:OPERation
The STATus:OPERation register provides an overview of state transitions of the tasks
(e.g. GPRF:MEASurement1:POWer) by collecting the information of lower registers. The
STATus:OPERation register hierarchy is shown below. The paths can also be queried
via remote command, see SYSTem:HELP:STATus[:REGister]? on page 259 and
SYSTem:HELP:STATus:BITS? on page 258.
Status Reporting System