Command Reference - Base Software
R&S
®
CMW500
223User Manual 1173.9463.02 ─ 06
Parameters:
'<FileDestination>' String parameter specifying path and filename of the target file.
Wildcards are not allowed.
Example:
MMEM:SAV '@SAVE\mysavefile.dfl'
Saves the current instrument settings to the file
mysavefile.dfl located in the directory assigned to the
@SAVE alias.
Usage: Event
Firmware/Software: V2.0.10
MMEMory:
STORe:ITEM '<ItemName>', '<FileDestination>'
Executes a partial save, i.e. stores the current settings of a specific application to the
specified file.
Parameters:
'<ItemName>' String parameter identifying the application to be saved
Examples: "GPRF Meas2", "WCDMA Gen1", "LTE Sig1"
'<FileDestination>' String parameter specifying path and filename of the target file.
Wildcards are not allowed.
Example:
MMEM:STOR:ITEM 'GPRF Meas1',
'@SAVE\mysavefile.dfl'
Saves the settings of the GPRF measurement instance 1 to the
file mysavefile.dfl located in the directory assigned to the
@SAVE alias.
Usage: Event
Firmware/Software: V3.0.10
MMEMory:STORe:STATe <MemoryNumber>, '<FileDestination>'
Stores the instrument settings from the specified internal memory to the specified file.
To store the current instrument settings to a file, use first *SAV <MemoryNumber> to
store the settings to the memory and then this command to store the settings from the
memory to a file.
For more convenience see MMEMory:SAV on page 222.
Parameters:
<MemoryNumber> Number of the internal memory to which the settings have been
stored using *SAV
'<FileDestination>' String parameter specifying path and filename of the target file.
Wildcards are not allowed.
Instrument-Control Commands