Command Reference - Base Software
R&S
®
CMW500
262User Manual 1173.9463.02 ─ 06
STATus:OPERation:TASK:A:<netw_std>:<func_grp>:PTRansition <FilterBits>
STATus:OPERation:TASK:A:<netw_std>:<func_grp>:<appl>:PTRansition
<FilterBits>
Sets the positive transition filter. If a bit is set, a 0 to 1 transition in the corresponding bit
of the associated condition register causes a 1 to be written in the associated bit of the
corresponding event register. See also Structure of an SCPI Status Register.
For a description of the variables <netw_std>, <func_grp> and <appl> refer to table 6-1.
Parameters:
<FilterBits> Range: 0 to 65535 (decimal representation)
Example:
STAT:OPER:PTR 1536
Set bits no. 9 and 10 of the PTRansition part for STATus:OPER-
ation register (1536 = 512 + 1024 = 2
9
+ 2
10
).
Firmware/Software: V1.0.4.11
STATus:OPERation:WCONdition? <WaitBits>[, <Timeout>]
STATus:OPERation:TASK:WCONdition? <WaitBits>[, <Timeout>]
STATus:OPERation:TASK:A:WCONdition? <WaitBits>[, <Timeout>]
STATus:OPERation:TASK:A:<netw_std>:WCONdition? <WaitBits>[, <Timeout>]
STATus:OPERation:TASK:A:<netw_std>:<func_grp>:WCONdition? <WaitBits>[,
<Timeout>]
STATus:OPERation:TASK:A:<netw_std>:<func_grp>:<appl>:WCONdition?
<WaitBits>[, <Timeout>]
Waits until at least one bit is set to true in the CONDition part of the status register, that
has also been set in the <WaitBits>. In other words, waits until an AND operation between
the bit patterns yields a positive result.
An optional timeout can be defined. When the AND operation yields a positive result or
the timeout is reached, the command returns the result of the AND operation.
The command can be used e.g. to wait until a measurement has been finished before
querying the measurement results.
For a description of the variables <netw_std>, <func_grp> and <appl> refer to table 6-1.
Query parameters:
<WaitBits> Range: 0 to 65535 (decimal representation)
<Timeout> Timeout in ms
Return values:
<Result> Result of the AND operation. A 0 indicates that a timeout occured.
Range: 0 to 65535 (decimal representation)
Instrument-Control Commands