Command Reference - Base Software
R&S
®
CMW500
261User Manual 1173.9463.02 ─ 06
STATus:OPERation[:EVENt]?
STATus:OPERation:TASK[:EVENt]?
STATus:OPERation:TASK:A[:EVENt]?
STATus:OPERation:TASK:A:<netw_std>[:EVENt]?
STATus:OPERation:TASK:A:<netw_std>:<func_grp>[:EVENt]?
STATus:OPERation:TASK:A:<netw_std>:<func_grp>:<appl>[:EVENt]?
Returns the contents of the EVENt part of the status register. Reading an EVENt part
clears it. See also Structure of an SCPI Status Register.
For a description of the variables <netw_std>, <func_grp> and <appl> refer to table 6-1.
Return values:
<EventBits> Range: 0 to 65535 (decimal representation)
Example:
STAT:OPER:TASK:A:GPRF:MEAS:POW?
Query the EVENt part of the status register for the GPRF power
measurement to check whether an event has occurred since the
last reading.
Usage: Query only
Firmware/Software: V1.0.4.11
STATus:OPERation:NTRansition <FilterBits>
STATus:OPERation:TASK:NTRansition <FilterBits>
STATus:OPERation:TASK:A:NTRansition <FilterBits>
STATus:OPERation:TASK:A:<netw_std>:NTRansition <FilterBits>
STATus:OPERation:TASK:A:<netw_std>:<func_grp>:NTRansition <FilterBits>
STATus:OPERation:TASK:A:<netw_std>:<func_grp>:<appl>:NTRansition
<FilterBits>
Sets the negative transition filter. If a bit is set, a 1 to 0 transition in the corresponding bit
of the associated condition register causes a 1 to be written in the associated bit of the
corresponding event register. See also Structure of an SCPI Status Register.
For a description of the variables <netw_std>, <func_grp> and <appl> refer to table 6-1.
Parameters:
<FilterBits> Range: 0 to 65535 (decimal representation)
Example:
STAT:OPER:TASK:A:GPRF:NTR 1536
Set bits no. 9 and 10 of the NTRansition part of the GPRF status
register (1536 = 512 + 1024 = 2
9
+ 2
10
).
Firmware/Software: V1.0.4.11
STATus:OPERation:PTRansition <FilterBits>
STATus:OPERation:TASK:PTRansition <FilterBits>
STATus:OPERation:TASK:A:PTRansition <FilterBits>
STATus:OPERation:TASK:A:<netw_std>:PTRansition <FilterBits>