Command Reference - Base Software
R&S
®
CMW500
282User Manual 1173.9463.02 ─ 06
Example:
SYSTem:CONNector:TRANslation? RFAC
Returns e.g. RFAC, RF1C
SYSTem:CONNector:TRANslation? RF2C
Returns e.g. RFBC, RF2C
Usage: Query only
Firmware/Software: V2.0.10
FORMat:BASE[:DATA] <Format>, [<Length>]
Selects the format for numeric data transferred to and from the R&S CMW500.
The format setting is only valid for commands and queries whose description states that
the response is formatted as described by FORMat:BASE:DATA. In particular, it affects
the results of the GPRF "I/Q Recorder" measurement.
Parameters:
<Format> ASCii | REAL
ASCii: Numeric data is transferred as ASCII bytes. The numbers
are separated by commas as specified in IEEE 488.2.
REAL: Data is transferred in a definite length block as IEEE float-
ing point numbers of the specified <Length>. See Block Data For-
mat.
<Length> This optional parameter is needed for REAL format only. It defines
the length of the floating point numbers in bits. The R&S CMW500
supports 32-bit (4-byte) numbers.
Usage: SCPI confirmed
Firmware/Software: V1.0.5.3
SYSTem:BASE:RELiability?
Returns a reliability value indicating errors detected by the base software.
Return values:
<Value>
For reliability indicator values, see chapter 5.4.4.1, "Reliability
Indicator", on page 149
Usage: Query only
Firmware/Software: V2.0.10
SYSTem:GENerator:ALL:OFF
SYSTem:MEASurement:ALL:OFF
SYSTem:SIGNaling:ALL:OFF
Switch off all generators, measurements or signaling applications.
Instrument-Control Commands