DISPlay
. :LEVel
. . :RANGe? [MINimum|MAXimum]
Query range of signal levels that is displayed in a level-bar view
This setting only applies if the tone mode if off (OUTP:TONE:STAT OFF). If
tone mode is on, the device determines the lower limit itself.
Parameter:
none query of current range
MINimum|MAXimum query of minimum|maximum range
Result:
signal-level range in dBZV
Example:
DISPlay:LEVel:RANGe? -> 30
DISPlay
. :PSCan
. . :LEVel
. . . :RANGe <numeric_value>|MINimum|MAXimum
Sets the range of signal levels that is displayed in a panorama view. Different
levels within this range can be distinguished in the view. The top-end of the
range is equal to “DISP:PSCAN:LEV:REF”.
Parameters:
<numeric_value> signal-level range in dBZV
MINimum|MAXimum minimum|maximum range
Example:
DISPlay:PSCAN:LEVel:RANGe 30