R&S PR100 User Manual
[SENSe]
. :FREQuency
. . :CONVersion
. . . :THReshold <numeric_value>|MINimum|MAXimum
Set the conversion threshold. This determines at which frequency the device
switches from normal to direct path reception. During normal reception, the
signal is modulated downward to an intermediate frequency. When the
frequency drops below the conversion threshold, the downward modulation is
skipped, hence the name “direct path” reception.
Parameters:
<numeric_value> value of conversion threshold
MINimum|MAXimum setting the lowest|highest conversion threshold
Example:
SENSe:FREQuency:CONVersion:THReshold 27 MHz
[SENSe]
. :FREQuency
. . :CONVersion
. . . :THReshold? [MINimum|MAXimum]
Query of conversion threshold.
Parameters:
none query of current conversion threshold
MINimum|MAXimum query of lowest|highest conversion threshold
Result:
conversion threshold in Hz
Example:
SENSe:FREQuency:CONVersion:THReshold? -> 27000000