[SENSe]
. :FREQuency
. . :PSCan
. . . :STOP <numeric_value>|MINimum|MAXimum
Setting the stop frequency of an RF-panorama scan. This value is mapped
onto the same variable as the frequency scan stop frequency, meaning that if
the stop frequency changes for PSCan it automatically changes for frequency
scan, SWE, as well. This setting modifies SENS:FREQ:PSC:SPAN and
SENS:FREQ:PSC:CENT.
The start frequency must be smaller than the stop frequency. A start
frequency that is larger than the stop frequency is rejected with error -
221(“Settings conflict”).
Parameters:
<numeric_value> frequency
MINimum|MAXimum setting the lowest|highest stop frequency
Example:
FREQuency:STOP 136 MHz
[SENSe]
. :FREQuency
. . :PSCan
. . . :STOP? [MINimum|MAXimum]
Query of a stop frequency of an RF-panorama scan. This command is alias
for SENS:FREQ:STOP?
Parameters:
none query of current stop frequency
MINimum|MAXimum query of lowest|highest stop frequency
Result:
Frequency in Hz
Example:
FREQuency:STOP? -> 136000000