R&S PR100 User Manual
[SENSe]
. :FREQuency
. . :PSCan
. . . :STARt <numeric_value>|MINimum|MAXimum
Setting of start frequency of an RF-panorama scan. This value is mapped
onto the same variable as the frequency scan start frequency, meaning that if
the start frequency changes for PSCan it automatically changes for frequency
scan, SWE, as well. This setting modifies SENS:FREQ:PSC:SPAN and
SENS:FREQ:PSC:CENT.
The start frequency must be smaller than the stop frequency. A start
frequency that is larger than the stop frequency is rejected with error -
221(“Settings conflict”).
Parameters:
<numeric_value> frequency
MINimum|MAXimum setting the lowest|highest start frequency
Example:
FREQuency:STARt 118 MHz
[SENSe]
. :FREQuency
. . :PSCan
. . . :STARt? [MINimum|MAXimum]
Query of start frequency of an RF-panorama scan.This command is alias for
SENS:FREQ:STAR?.
Parameters:
none query of current start frequency
MINimum|MAXimum query of lowest|highest start frequency
Result:
Frequency in Hz
Example:
FREQuency:STARt? -> 118000000