R&S PR100 User Manual
[SENSe]
. :FREQuency
. . :STARt <numeric_value>|MINimum|MAXimum
Setting of start frequency of a frequency scan. This value is mapped onto the
same variable as the PSCan start frequency, meaning that if the start
frequency changes for frequency scan, SWE, it automatically changes for
PSCan as well. The start frequency must be smaller than the stop frequency.
A start frequency that is larger than the stop frequency is rejected with error -
221(“Settings conflict”).
Parameters:
<numeric_value> frequency
MINimum|MAXimum setting the lowest|highest start frequency
Example:
FREQuency:STARt 118 MHz
[SENSe]
. :FREQuency
. . :STARt? [MINimum|MAXimum]
Query of start frequency of a frequency scan.
Parameters:
none query of current start frequency
MINimum|MAXimum query of lowest|highest start frequency
Result:
Frequency in Hz
Example:
FREQuency:STARt? -> 118000000