EasyManua.ls Logo

Rohde & Schwarz PR100 - Page 191

Rohde & Schwarz PR100
282 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
R&S PR100 User Manual
[SENSe]
. :FREQuency
. . :STOP <numeric_value>|MINimum|MAXimum
Setting the stop frequency of a frequency scan. This value is mapped onto the
same variable as the PSCan stop frequency, meaning that if the stop
frequency changes for frequency scan, SWE, it automatically changes for
PSCan as well. The start frequency must be smaller than the stop frequency.
A start frequency that is larger than the stop frequency is rejected with error -
221(“Settings conflict”).
Parameters:
<numeric_value> frequency
MINimum|MAXimum setting the lowest|highest stop frequency
Example:
FREQuency:STOP 136 MHz
[SENSe]
. :FREQuency
. . :STOP? [MINimum|MAXimum]
Query of a stop frequency of a frequency scan.
Parameters:
none query of current stop frequency
MINimum|MAXimum query of lowest|highest stop frequency
Result:
Frequency in Hz
Example:
FREQuency:STOP? -> 136000000

Table of Contents

Related product manuals