R&S PR100 User Manual
MTRACE
Output of measured values of all sensor functions switched on. If no function
is switched on, NaN (Not a Number) is output. The INF value 9.9E37 is
entered into the result buffers MTRACE and ITRACE to mark the end of the
trace.
1. “VOLTage:AC" (if the associated function is switched on via
SENS:FUNC:ON)
2. "FREQuency:OFFSet" (if the associated function is switched on via
SENS:FUNC:ON)
3. “FSTRength" (if the associated function is switched on via
SENS:FUNC:ON)
The end of each sweep (if SENS:SWE:CNT is set to 10, then there are 10
sweeps in a scan) is marked by the values 2000 for “VOLTage:AC" and 0 for
Frequency.
ITRACE
1. Channel Number
2. Frequency
The end of each sweep (if SENS:SWE:CNT is set to 10, then there are 10
sweeps in a scan) is marked by the value 0 for Frequency.