Functions
8.6 Configuration
FIDAMAT 6
108 Operating Instructions, 01/2019, A5E00222135-04
Measuring range 1 to 4 Measuring range 1 to
Measuring range identifica-
Limit 1 to 4 Limit 1 (to 4) triggered Limit alarm on/off
Control function
(CTRL)
Function control active
(e.g. on change to
operator control
mode;
in the warm-up phase,
when AUTOCAL or
calibration is running)
Function control
deactivated
Signal when:
• Device is decoded
• Warm-up phase
(max. 2 h)
• Calibration running
(AUTOCAL)
• Remote control via inter-
face (ELAN,
PROFIBUS)
Sample gas Supply of sample gas Actuation of valves
Calibration gas 1 to 4 Supply of calibration
Actuation of valves
Measuring point 1 to 6 Measuring point 1 to
6 selected
For taking a gas sample via
solenoid valves at various
Signal for measuring
point 1 to 6
Measuring point 1 to
6 selected
For identifying the sample
point (runs in parallel with
Signaling contact When signaling, the
relay is briefly ener-
gized.
e.g. in the case of
AUTOCAL:
Control of a second device
Gas path heating Heating ready for
For identification purposes
Autocal check Autocal difference
too large (func-
Combustion gas Combustion gas
Actuation of valves
Combustion air Combustion air valve
Actuation of valves
Operator input prompt Operator intervention
required
Gas connection faulty;
W5 (combustion
air/combustion gas) pre-
sent;
Device in PAUSE /