Service and maintenance
9.2 General information on maintenance
FIDAMAT 6
Operating Instructions, 01/2019, A5E00222135-04
135
NOTICE
Penetration of moisture into the device
Device damage.
Make sure when carrying out cleaning and maintenance work that no moisture
penetrates the inside of the device.
9.2 General information on maintenance
Depending on the use of the device and certain empirical values, determine a maintenance
interval for the tests to be carried out repeatedly.
The maintenance intervals will vary from site to site depending on corrosion resistance.
9.3 Maintaining the device
Perform the following maintenance work:
Check the device annually:
For electrical safety and function
For leakage of the sample gas path within the device
Check the device for leaks after every intervention in the gas path. Refer to section
"Checking for leaks (Page 56)" for information on this.
FIDAMAT 6-E: Perform pump maintenance every 6 months. Replace the pump
diaphragm.
Note
If gaskets that come in contact with sample gas are not expected to be negatively affected
by chemical corrosion, the owner may at its discretion extend the maintenance interv
al in
special cases.