Functions
8.6 Configuration
FIDAMAT 6
Operating Instructions, 01/2019, A5E00222135-04
127
8.6.17 Error on/off (function 87)
Figure 8-63 Switching errors on and off
With this function, you switch the message for maintenance requests and faults off
individually so that there is no entry in the logbook, no message in the status line of the
measured value display and no external signal sent.
8.6.18 PROFIBUS configuration (function 90)
This function is used to set the PROFIBUS parameters. This function can only be called if
the device contains additional PROFIBUS electronics (optional board).
Figure 8-64 Configuring PROFIBUS
Specifically, you can set the following specific PROFIBUS parameters:
● Address (PROFIBUS station address): The address can be set from 0 to 126.
● Under TAG you can display an identifier assigned to the device in the PROFIBUS
network (or the first 16 characters thereof).
● You set the configuration behavior of the device with the
"Ident number"
(softkey 2).
Parameters 0, 1 and 3 can be selected and have the following meaning: