Functions
8.6 Configuration
FIDAMAT 6
116 Operating Instructions, 01/2019, A5E00222135-04
8.6.6 Reset (function 74)
Figure 8-48 Carrying out a reset
With this function, you can carry out a warm restart.
After calling this function, the device switches to the "Warm-up phase", which activates the
functional check at the same time. The device is only completely ready for operation again
after successfully running through this phase.
8.6.7 Save / load data (function 75)
Figure 8-49 User data memory
You configure the user data memory with this function.
Press the softkey 1
"Save user data"
e.g. after the system has been successfully
commissioned. All individual settings are then saved.
Press the softkey 2
"Load user data"
, to load the last user data saved.
These functions are important when the device is undergoing repairs or maintenance work,
or, for example, if new parameter settings are to be tried out.
Press softkey 3 to restore the factory settings.