Commissioning
6.3 Commissioning and operation
FIDAMAT 6
62 Operating Instructions, 01/2019, A5E00222135-04
6.3.3 State change due to fault / maintenance request
In the following tables, you will find possible automatic state changes of the device:
Table 6- 1 Change of state due to fault
S2: Pump does not start up
(FIDAMAT 6-E only)
Warming-up | Standby Standby
S3: Flame does not ignite
S4: External fault - - -
S5: Oven temperature Warming-up | Standby | Measuring
Pause without heating
S7: Flame temperature Warming-up | Standby | Measuring
Pause
S8: Combustion air/combustion
Warming-up | Standby | Measuring
Pause
S10: 24 h RAM/Flash check
S12: Line voltage supply Warming-up | Standby | Measuring
Pause
S13: Hardware Warming-up | Standby | Measuring
Pause
S14: Measured value > Full-scale
- - -
Table 6- 2 Change of state due to maintenance request
W1: Calibration difference too
- - -
W4: Set clock - - -
W5: Combustion air/combustion
Standby | Measuring Warming-up
W6: Temperature sensor display - - -
W7: Electronics/hardware tempera-
- - -
W9: External maintenance request
W10: AUTOCAL check deviation
After you have eliminated the cause of a maintenance request or fault, exit the current
operating state using function 91.