Functions
8.6 Configuration
FIDAMAT 6
126 Operating Instructions, 01/2019, A5E00222135-04
8.6.16 Linear temperature compensation (function 86)
Figure 8-62 Compensating the temperature
The FIDAMAT 6 is temperature compensated both at the zero point as well as for the span.
If an additional temperature error occurs during operation, compensate it with this function.
"Aftercompensation of the zero point":
Based on an average analyzer temperature, define two different correction values for ranges
of increased temperature decreased temperature.
"Aftercompensation of the span":
The procedure is the same as for the zero point, but the change relates to the measured
value.
Note
If the zero point deviates negatively when the
temperature changes, the value has a positive
-versa. The same applies to the measured value.