Functions
8.6 Configuration
FIDAMAT 6
130 Operating Instructions, 01/2019, A5E00222135-04
8.6.20 Pressure values (function 92)
Figure 8-67 Changing the pressure values
Different pressure values are displayed in this function. You can change some of these
values.
Combustion gas pressure:
Is automatically displayed
Max. difference If the combustion gas pressure deviates from the
setpoint pressure by this value, the device exits
the "Measuring" operating state. This value can
Actual pressure Is automatically displayed
Pressure combustion air:
Is automatically displayed
Max. difference If the combustion air pressure deviates from the
setpoint pressure by this value, the device exits
the "Measuring" operating state. This value can
Min. rise for starting pump When the pump starts, the combustion air pres-
sure must rise by this value for the device to go to
"Measuring" operating state. This value can be
Is automatically displayed