Technical specifications
11.2 Technical specifications of the FIDAMAT 6-E
FIDAMAT 6
Operating Instructions, 01/2019, A5E00222135-04
169
Table 11- 8 Technical specifications Climatic conditions
Climatic conditions
Permissible ambient temperature +5 °C to +45 °C (41 °F to 113 °F) during operation as a single device
+5 °C to +40 °C (41 °F to 104 °F) during operation in analysis cabinets
-30 °C to +70 °C (-22 °F to 158 °F) during storage and transport;
Permissible humidity < 90% relative humidity annual average,
during storage and shipping (dropping below dew point not permissi-
ble)
11.2 Technical specifications of the FIDAMAT 6-E
Table 11- 9 Technical specifications General
General
Measuring ranges 4, internally and externally switchable;
manual and automatic measuring range switchover possible
Automatic measuring range switchover
Hysteresis, selectable
Smallest possible span
0 ... 10 ppm
Detection limit
0.1 ppm
Largest possible span
99999 ppm
Concentration units
ppm C1, ppm C3, ppm C6, mg C/m
3
Measured value display
Digital concentration display (5 digits with floating decimal point)
Resolution of digital display
0.1% of measured value
Operating position
Front panel vertical
Oven temperature Can be set between 100 and 200 °C (212 and 392 °F), application-
dependent
Table 11- 10 Technical specifications Gas inlet conditions
Gas inlet conditions
Sample gas pressure
600 ... 1100 hPa
Sample gas flow rate
Approx. 1 l/min (60 l/h)
Sample gas temperature
0 ... 200 °C (32 ... 392 °F)
Sample gas humidity
< 90% relative humidity